Electrical Insulation from Moxa
You have been filtering for Moxa and therefore you see now only Electrical Insulation from Moxa displayed.
You can choose other manufacturers on the left side in the menu or get the complete overview.
Electrical Insulation from Moxa
|
Moxa TCC-82 , optischer Isolator für RS232
4-channel 4 KV RMS Isolation for 1 minuteDual self-powered or selectable external power sourceSurge protection, 15 KV ESDAuto baudrate sensingCompact sizeTCC-82 is a signal isolator that provides ... Lieske Part No.: 851508 Mfg Part No: TCC-82 |
78,00 EUR
![]() ≈ £66.61 | ≈ $88.76
Delivery time on requestinfo_outline
|
⚡ CPE – Central Potential Equalisation in Medical Locations
|
Mandatory per IEC 60364-7-710 / DIN VDE 0100-710: In Group 1 and Group 2 medical locations, a Central Protective Equipotential Bonding (CPE) system is mandatory. All conductive parts – enclosures, beds, infusion stands, pipework – are connected to a common bonding conductor (max. 0.1 Ωresistance) and protective earth. Purpose: fault currents cannot flow through the patient.
Group 1(X-ray, physiotherapy): leakage < 500 µA · Group 2(OR, ICU, cardiac cath lab): leakage < 10 µA · Inspection intervals: STKannually by qualified electrician, MTKevery 2 years by authorised inspector. |
⚡ Application Areas for Electrical Isolation Components
|
OR & Intensive Care
Group-2 locations: IT networks with monitoring, 2-MOPP power supplies, CPE with < 10 µA leakage. |
Imaging & Diagnostics
X-ray, CT, MRI: Group-1 requirements, leakage current limitation, insulation testing per IEC 60601. |
GP Practice & Outpatient
Minor procedures, ECG, ultrasound: isolation between mains and patient monitoring systems. |
|
Laboratory & IVD
Analysers, PCR systems: galvanic isolation prevents measurement interference from earth loops. |
OEM Medical Device Mfg.
Isolation components as device sub-assembly: certification evidence per EU MDR and IEC 60601-1. |
Related Categories
|
Medical Power Supplies
IEC-60601-1 PSUs (2 MOPP/1 MOPP) – complement isolation components in the complete system. |
Insulation Testers
Test equipment for STK/MTK inspections: leakage current, insulation resistance, protective earth per IEC 62353. |
Medical PC Components
IEC-60601-compatible industrial hardware for clinical systems and embedded medical devices. |
|
Medical Technology – Overview
All product areas for medical use at a glance. |
|
MOPP Reference
< 300 µA: MOOP (operator) · < 100 µA: 1 MOPP · < 10 µA: 2 MOPP (direct patient contact). PSUs and isolation must match the required protection level. |
OEM & Project Orders
Framework agreements, component reservation, and certification documentation on request for device manufacturers. |
Technical Advice
+49 (0)7666 / 88499-0 |