Electrical Insulation from Moxa
You have been filtering for Moxa and therefore you see now only Electrical Insulation from Moxa .
You can choose other manufacturers on the left side in the menu or get the complete overview.
Electrical Insulation from Moxa
|
Moxa TCC-82 , optischer Isolator für RS232
4-channel 4 KV RMS Isolation for 1 minuteDual self-powered or selectable external power sourceSurge protection, 15 KV ESDAuto baudrate sensingCompact sizeTCC-82 is a signal isolator that provides ... Lieske Part No.: 851508 Mfg Part No: TCC-82 |
78,00 EUR
![]() ≈ £67.39 | ≈ $90.48 plus delivery charges
Delivery time on requestinfo_outline
|
⚡ CPE – Central Potential Equalisation in Medical Locations
|
Mandatory per IEC 60364-7-710 / DIN VDE 0100-710: In Group 1 and Group 2 medical locations, a Central Protective Equipotential Bonding (CPE) system is mandatory. All conductive parts – enclosures, beds, infusion stands, pipework – are connected to a common bonding conductor (max. 0.1 Ωresistance) and protective earth. Purpose: fault currents cannot flow through the patient.
Group 1(X-ray, physiotherapy): leakage < 500 µA · Group 2(OR, ICU, cardiac cath lab): leakage < 10 µA · Inspection intervals: STKannually by qualified electrician, MTKevery 2 years by authorised inspector. |
⚡ Application Areas for Electrical Isolation Components
|
OR & Intensive Care
Group-2 locations: IT networks with monitoring, 2-MOPP power supplies, CPE with < 10 µA leakage. |
Imaging & Diagnostics
X-ray, CT, MRI: Group-1 requirements, leakage current limitation, insulation testing per IEC 60601. |
GP Practice & Outpatient
Minor procedures, ECG, ultrasound: isolation between mains and patient monitoring systems. |
|
Laboratory & IVD
Analysers, PCR systems: galvanic isolation prevents measurement interference from earth loops. |
OEM Medical Device Mfg.
Isolation components as device sub-assembly: certification evidence per EU MDR and IEC 60601-1. |
Related Categories
|
Medical Power Supplies
IEC-60601-1 PSUs (2 MOPP/1 MOPP) – complement isolation components in the complete system. |
Insulation Testers
Test equipment for STK/MTK inspections: leakage current, insulation resistance, protective earth per IEC 62353. |
Medical PC Components
IEC-60601-compatible industrial hardware for clinical systems and embedded medical devices. |
|
Medical Technology – Overview
All product areas for medical use at a glance. |
|
MOPP Reference
< 300 µA: MOOP (operator) · < 100 µA: 1 MOPP · < 10 µA: 2 MOPP (direct patient contact). PSUs and isolation must match the required protection level. |
OEM & Project Orders
Framework agreements, component reservation, and certification documentation on request for device manufacturers. |
Technical Advice