| Additional features |
| Alarm threshold setting | Alarm: 50 - 500 kOhm |
| Measured value type | Network insulation resistance, range: 1 kOhm - 10 MOhm, precision: +/- 5 %; Alarm history |
| Configurable filter time | 1 µs |
| [tA] Response time | 0.8 s 1/5th of set filtering time; 8 s 1/5th of set filtering time; 32 s 1/5th of set filtering time |
| Time delay range | 0 - 120 min for alarm |
| With alarm function | IT transformer overheat; IT transformer overload |
| Self-test | Manual; Automatic test |
| Over voltage category | III |
| Internal impedance | 141 kOhm at 2.5 Hz |
| Degree of protection (IP) | Front: IP52; Rear: IP20 |
| Material | Thermoplastic |
| Communication & Management |
| Display type | LCD display |
| Displayed information | Resistance value; IT transformer alarms; Evolution of resistance in graph |
| User language | Russian; Spanish; French; Chinese; German; Portuguese; English; Italian |
| Local signalling | Green indicator light: satisfactory insulation resistance; Orange audible and visible alarms: below disturbance threshold |
| Output contact | 2 changeover contacts standard or fail-safe |
| Breaking capacity | 12 - 48 V - DC at 50 mA |
| Communications protocol | - |
| Data transmission | HMI: Test reports; Dry contact: Test reports |
| Fault event log | Alarm logs |
| Contractual warranty |
| Warranty (in months) | 18 |
| Dimensions |
| Installation type | Indoor |
| Relative humidity | 0…92 % |
| Operating ambient temperature | -25…55 °C |
| Ambient temperature during storage | -40…70 °C |
| Pollution degree | 2 |
| Mounting mode | DIN rail mountable; Flush mount (plate) |
| Connections - terminals | Removable screw terminal block 0.2…2.5 mm² |
| Product weight | 0,25 kg |
| Cutout dimensions | 92.4 x 92.4 mm |
| Height | 110 mm |
| Width | 96 mm |
| Depth | 59 mm |
| Main features |
| Product series | Vigilohm |
| Device short name | IM15H |
| Product or component type | Insulation monitoring device |
| Type of IMD | AC IMD |
| Earthing system | IT |
| Application | Group II medical premise; Healthcare |
| Circuit type | Control circuit |
| Network capacitance | Lightly capacitive |
| Type of network | 1P + N |
| Supply voltage | 0…230 V - AC at 50/60 Hz (device connected to neutral conductor) ; 0…230 V - DC |
| Packaging units |
| VPE 1 type | PCE |
| Number of devices per package | 1 |
| VPE 1 height | 10,5 cm |
| VPE 1 width | 13,5 cm |
| VPE 1 length | 16,5 cm |
| Packaging weight (Lbs) | 467,0 g |
| VPE 2 type | S03 |
| VPE 2 quantity | 8 |
| VPE 2 height | 30,0 cm |
| VPE 2 width | 30,0 cm |
| VPE 2 length | 40,0 cm |
| VPE 2 weight | 4,323 kg |
| Performance requirements |
| [UH,nom] Rated operating voltage | 110 - 230 V AC 50/60 Hz tolerance: ± 15 %; 125-250 V DC tolerance: ± 15 % |
| Power consumption in VA | 12 VA |
| Over current protection | 1 A |
| Standards |
| Standards | IEC 61010-1; IEC 61557-8; IEC 60364-7-710; IEC 61326-2-4; UL 61010-1 |
| Product certifications | DNV; EAC; BV; CE; EU-RO MR TA |
| Environmental properties | Humid heat, out of operation corresponds to IEC 60068-2-30; humid heat, in operation corresponds to IEC 60068-2-56; salt mist corresponds to IEC 60068-2-52; long-term storage class 1K22 corresponds to IEC 60721-3-1; transport class 2K11 corresponds to IEC 60721-3-2; stationary use at weather protected locations class 3K24 corresponds to IEC 60721-3-3 |
| Mechanical robustness | Long-term storage class 1M11 corresponds to IEC 60721-3-1; Transport class 2M4 corresponds to IEC 60721-3-2; Stationary use class 3M11 corresponds to IEC 60721-3-3 |
| Supported hardware |
| Product compatibility | Automatic fault locator |
Dieses Gerät gehört zur Vigilohm-Reihe, die für die Überwachung von ungeerdeten/IT-Netzen gemäß den Installations- und Produktnormen entwickelt wurde. Nach den IEC-Normen ist die Ausstattung von medizinischen IT-Räumen mit Isolationsüberwachungsgeräten (IMD) vorgeschrieben. Dieses Gerät ist für die Überwachung von AC-, DC- und ACDC-Systemen der medizinischen IT bestimmt. Er misst den Isolationswiderstand des Systems durch Einspeisung eines niederfrequenten Wechselstroms zwischen dem Netz und der Erde. Die Konfiguration wird in Übereinstimmung mit den medizinischen Standards von der Vorderseite des Geräts aus über den HMI-Bildschirm vorgenommen. Die Einstellungen beziehen sich auf Isolationsalarmschwellen von 50 Kilohm bis 500 Kilohm und Transformatorüberhitzungs- und Überlastalarme. Es überwacht Netze bis zu 230VAC, 230VDC, in der medizinischen IT-Architektur, sowie den medizinischen IT-Transformator. Der schwarz-weiße Bildschirm kann einen Isolationswiderstand von 100 Ohm bis 10 Megohm anzeigen. Der Autotest startet automatisch alle 5 Stunden und bei Bedarf. Fällt der Isolationswiderstand des Systems unter den eingestellten Schwellenwert, d.h. es liegt ein Erdschluss vor, wird über das IMD-Ausgangsrelais innerhalb von 4 Sekunden ein Alarm ausgelöst. Sobald kein Fehler mehr überwacht wird, wird der Alarm gelöscht, das Alarmrelais ist wieder normal. Das Gerät kann von einer 110VAC bis 230VAC Quelle oder einer 125VDC bis 250VDC Quelle versorgt werden, alle mit einer Toleranz von plus oder minus 15%. Dieses Produkt kann auf einer DIN-Schiene oder bündig installiert werden. Die Vorderseite hat die Schutzart IP52, die Rückseite IP20 und Überspannungskategorie III. Das Überwachungsgerät ist mit einem RS485-Kommunikationsport, Modbus-Protokoll, ausgestattet.