Further information for ifm Electronic PH1706 pressure sensor -0,13...2,5 bar relative pressure
Accuracy / Deviations
Repeatability
< ± 0.1; (at temperature variations < 10 K; Turn down 1:1) % of span
Characteristic deviation
< ± 0.2; (linearity including hysteresis and repeatability, end-point setting according to DIN EN IEC 62828-1) % of span
Linearity deviation
< ± 0.15; (turn down 1:1) % of span
Hysteresis deviation
< ± 0.15; (turn down 1:1) % of span
Long-term stability
< ± 0.1; (turn down 1:1; per year) % of span
Total deviation over temperature range
Temperature range / total deviation; -25...15 °C / characteristic deviation ± 0.05 % of span / 10 K; 15...80 °C / characteristic deviation; 80...150 °C / characteristic deviation ± 0.1 % of span / 10 K
Notes on accuracy / deviation
Further information see section Diagrams and curves.
Approvals / tests
EMC
DIN EN 61000-6-2; DIN EN 61000-6-3
Shock resistance
DIN EN 60068-2-27 / 50 g (11 ms)
Vibration resistance
DIN EN 60068-2-6 / 20 g (10...2000 Hz)
MTTF
323 years
Electrical connection
Plug connection
1 x M12; Coding: A; Contacts: 4, gold-plated
Electrical data
Operating voltage
18...30 DC V
Minimum insulation resistance
100; (500 V DC) MΩ
Protection class
III
Electricity consumption
3,5...21,5 mA
Standby delay time
1 s
Reverse polarity protection
Integrated watchdog
Environmental conditions
Ambient temperature
-25...80 °C
Storage temperature
-40...100 °C
Degree of protection
IP 67; IP 68; IP 69K
Field of application
Special feature
Gold-plated contacts
Model
PH1706: ZZPY1706+E30124+E33308
Measuring element
Ceramic-capacitive pressure sensor cell
Application
Flush-mounted for food and beverage industry
Media
Paste and solid-containing media; Liquid and gaseous media
Medium temperature
-25...150 °C
Min. bursting pressure
50000 mbar / 725 psi / 5000 kPa
Compressive strength
20000 mbar / 290 psi / 2000 kPa
Vacuum tightness
-1000 mbar
Pressure type
Gauge pressure
MAWP for applications according to CRN
20 bar
Backlash-free
Interfaces
Communication interface
IO-Link
Transmission type
COM2 (38.4 kBaud)
IO-Link revision
1.1
Profiles
Smart Sensor - SSP 3.1 / Measuring Sensor; Common - I&D / Identification and Diagnosis
-0.13...2.5 bar / -1.82...36.26 psi / -12.5...250 kPa
Process connection
Clamp DN40 DIN 11864-3-A-BKS
Response times
Attenuation analog output dAA
0...4 s
Step response time analog output
30 ms
Elektronischer Drucksensor, -0,13...2,5 bar, -1,82...36,26 psi, -12,5...250 kpa, Clamp DN40, Analogsignal, 4...20, (skalierbar) mA, Steckverbindung, -25...150 °C, IP 67, IP 68, IP 69K, Relativdruck