Further information for Baaske Medical 2006633 - MI 1005 galvanic isolation network installation
The technology is based on the MI 1005E 10,000-proven technology of the world Brought to use MI 1005. In contrast to the MI 1005 Medical insulator MI 1005E is suitable for installation in housing.
thesis are the basic components of HARTING Germany GmbH & Co . KG are available. thesis exported to IP65/67 push-pull interfaces are Standardized Accor ding to IEC 61076-3-106 (version 4).
1005E The MI isolates the Ethernet connection of a device with a voltage rating of up to 5,000 v , both the shield and the signal. Tested to medical standards, Malthus Allowing the MI 1005E Ethernet connection electrically safe medical electrical devices.
Suitable for international use and worldwide acceptance of MI 1005E what the unit of UL (Underwriters Laboratories), accor ding to UL 60601 - . 1 on its electrical and material properties tested withstand voltage (signal and shielding): 5000V 50/60 Hz for 1 min throughput: 10/100/1000 Mbit / s Standard: IEEE 802.3 standard tests: UL 60601-1 (E342310) RoHS and UL compliant materials V94 Dimensions: 51 x 23.4 x 18.2 mm
Weight: 20g
Connectors: 2 x RJ45
Material: UL94V-0 / RoHS-compliant
operating time between failures: 12:21 x 10 ^ 8