2007337

2007337

Medical Technology > Medical power supplies

2007337

Lieske Part No. 764555
Mfg Part No 2007337
EAN-Code
Manufacturer
Here only 10.798,92 EUR
12.850,71 EUR incl. VAT

Further information for 2007337

General
Nominal mains voltage230 V ~ 50/60 Hz
Operating range207 - 253 VAC
Trip value<= 50 kOhm
Response time< 2 sec.
Ambient operating temperature0 - 40 °C
Relative humidity30 - 75 %
Protection classII
Degree of protectionIP 40
Power consumption1,3 VA
StandardsEN60601-1, EN60601-1-1, EN60601-1-2, EN14971
The Isolationswächter MED R 3rd ISOG is a specialised monitoring sensor for medical power supply systems, designed for detection and alerting of insulation faults in safety-separated circuits. It is intended exclusively for use with isolation transformers of the MED R 3rd series and serves to ensure electrical safety in medical applications. Device monitoring is performed by continuous measurement of the insulation resistance between the two live contacts of the output socket of the isolation transformer. If the resistance falls below the defined threshold, an alarm is triggered.

Applications
This insulation monitor is designed for use in medical facilities where patients or operating personnel could be endangered by electrical insulation faults. Since isolation transformers provide galvanic separation between the mains and the patient circuit, upstream RCDs (residual current devices) cannot detect insulation faults. The insulation monitor closes this safety gap by detecting changes in insulation resistance that indicate a short circuit or insulation degradation.

Functions and Features in Detail
The monitor detects insulation faults as soon as the resistance drops below 50 kOhm. The response time is less than 2 seconds, enabling rapid reaction to hazardous situations. The alert is indicated both visually via an LED and acoustically via a beep tone, to draw the attention of operating personnel. An integrated function test can be performed using a test button to verify the device’s operational status.

Technology Explained Simply
The insulation monitor supervises the electrical resistance between the two conductors of the output socket of the isolation transformer. A sudden drop in resistance, for example due to a short circuit, is interpreted as an insulation fault. Monitoring is performed directly at one of the output sockets of the isolation transformer, without requiring additional wiring or external power supply. The operating voltage is 230 V ~ 50/60 Hz, with a tolerance range of ±10 %, ensuring stable operation in typical medical power supply systems. Device class II and protection rating IP 40 ensure reliable operation under normal indoor conditions.

Compatibility and Accessories
The insulation monitor is developed exclusively for operation with isolation transformers of the MED R 3rd series. It is not intended for use with other isolation transformers or power supply systems. Compatibility is limited to this specific product series to ensure optimal safety functions.

Mounting and Installation
Installation is performed by directly connecting to one of the output sockets of the isolation transformer. No additional mounting or connection components are required. The device dimensions are 125 x 67 x 40 mm (H x W x D), allowing for compact placement. The operating temperature range is between 0 °C and +40 °C, and the relative humidity must be between 30 % and 75 %.

Operational Advantages
Continuous monitoring of insulation resistance enhances electrical safety in medical applications. The clear visual and acoustic warning enables rapid response to potential hazards. The integrated test button ensures regular functional checks to maintain system reliability. Compliance with standards EN60601-1, EN60601-1-1, EN60601-1-2 and EN14971 ensures conformity with requirements of medical device legislation.
This is an offer of:

Lieske Elektronik e.K.

Reutener Str. 10
79279 Vörstetten

https://industry-electronics.com
Tel. 07666/88499-0
Fax. 07666/88499-111

Diesen Artikel finden Sie unter:
industry-electronics.com/artikel/764555
qrcode für 2007337