Further information for 2007190
MED R 1000VA isolation transformer 3rd 3rd Edition IEC 60601-1:2005, input voltage: 230 V, output voltage: 230 V, TÜV - approved, Number of Outputs: 9, including electronic Einschaltdämpfung, including mounting rails
Accor ding to the 60601-1 3rd Edition certified medical isolation transformer
Safety: EN 60601-1 (Medical electrical equipment)
EMC: EN 60601-1-2 (medical equipment) . The series of 3rd MED R corresponds to current medical standards and is tested accor ding to the 3rd edition The compact design Provides stability and flexible siting. Insensitive to electromagnetic changes in the medical separation Transformatorn the MED R 3rd series run smoothly and be Influenced by unwanted electrical or electromagnetic effects no other technical devices. With the medical isolation transformers MED R 3rd protect your patients from dangerous leakage currents Caused by voltage spikes and avoid failures of your equipment.
Specifications:
Rating: 1000 V
Input voltage: 230V
Output voltage: 230 V
Test voltage: 4 kV between primary and secondary windings
Number of Outputs: 9
Weight: 12,5 kg
Dimensions (HxWxD): 100 x 220 x 300 (+6 mm rubber feet)
Ambient temperature: max. 40 ° C
Storage temperature: -10 ° C to +50 ° C
Colour: RAL 7035 light gray
standards: IEC 60601-1 3rd Edition, EN 60601-1 3rd Edition, EN 60601-1-2, IEC 60601-1-2, UL 60601 . -1, CSA C22.2 601.1-M90, TUV - tested
including mounting rail.